Alarm Condition | Alarm Messages on LCD |
---|---|
Skin Sensor is unplugged /Faulty | Skin Sensor Fail |
Air Sensor is unplugged /Faulty | Air Sensor Fail |
Measured skin temperature higher than set temperature by 1.0°C | Skin Temp. High |
Measured skin temperature lower than set temperature by 1.0°C | Skin Temp. Low |
Measured skin temperature higher than 38.0°C (With cut-off) | Skin Over Range |
Measured skin temperature lower than 34.0°C | Skin Under Range |
Measured air temperature higher than set temperature by 1.5°C | Air Temp. High |
Measured air temperature lower than set temperature by 3.0°C | Air Temp. Low |
Measured air temperature higher than 39.0°C (With cut-off | Air Over Range |
Heater power is 100% for more than 15mins | Chk Sens Position |
Heater is unplugged or faulty | Heater Open |
Heater Power drive fail/uncontrollable | Chk. Pwr. Drive |
Power Fail Alarm | Audio with LED indicator |